Aptamer-Based Lateral Flow Biosensor for Rapid Detection of Salivary Cortisol.

ACS omega Dalirirad S, Han D, Steckl AJ
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

We have developed a disposable point-of-care (POC) aptamer-based biosensor for the detection of salivary cortisol. Nonstressful and noninvasive sampling of saliva compared to that of blood makes saliva an attractive biological matrix in developing POC devices for biomarker monitoring. Aptamers are attractive as recognition elements for multiple reasons, including their specific chemical synthesis, high stability, lack of immunogenicity, and cell-free evolution. A duplex aptamer conjugated to the surface of Au nanoparticles (AuNPs) by Au-S bonds is utilized as the sensor probe in a lateral flow assay (LFA) device. The addition of saliva samples containing cortisol makes the cortisol-aptamer undergo conformational changes and dissociate from the capture probe. Increasing cortisol concentration in the dispensed saliva sample results in increased dissociation and leads to increased binding of AuNP conjugate on the test line. Therefore, the color intensity of the test line on the LFA is a direct function of the concentration of cortisol in saliva. This simple and fast method provides detection in the cortisol range of ∼0.5-15 ng/mL, which is in the clinically accepted range for salivary cortisol. The limit of detection was 0.37 ng/mL, and the accuracy was confirmed by enzyme-linked immunosorbent assay (ELISA) testing results. High selectivity was observed for salivary cortisol against other closely related steroids and stress biomarkers present in saliva.