组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
In this study, we developed an electrochemical-based single-use neurobiosensor based on multiwalled carbon nanotube (MWCNT)-gold nanoparticle (AuNP) nanocomposite doped, 11-amino-1-undecanethiol (11-AUT)-modified polyethylene terephthalate coated indium tin oxide (ITO-PET) electrodes. This electrode was used for the sensitive determination of DJ-1, a protein responsible for mitochondrial dysfunction in Parkinson's disease (PD) with the task of eliminating oxidative stress. The design strategy and analytical studies for the neurobiosensor were monitored with electrochemical impedance spectroscopy (EIS), cyclic voltammetry (CV) and single frequency impedance (SFI) techniques. The selective determination range for DJ-1 of the developed neurobiosensor system is 4.7-4700 fg mL-1 in accordance with the charge transfer resistance (Rct) associated with a limit of detection of 0.5 fg mL-1. Since changes in the expression of DJ-1 protein is particularly important in cerebrospinal fluid (CSF) and saliva, the ability of the developed neurobiosensor system to detect the DJ-1 protein in these media was tested by the standard addition method. The statistical results show that the biosensor decorated with MWCNT-AuNP-AUT may be recommended for the selective determination of DJ-1 protein.