2020

Injectable Flexible Subcutaneous Electrode Array Technology for Electrocardiogram Monitoring Device.

ACS biomaterials science & engineering Bong J, Yasin O, Vaidya VR, Park J, Attia ZI, Padmanabhan D, Cho SJ, Asirvatham R, Schneider N, Lee J, Kim EM, Friedman PA, Ma Z
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Implantable cardiac monitors have undergone considerable miniaturization. However, they continue to be associated with complications such as infection, bleeding/bruising, and device extrusion or migration. In this paper, we demonstrate the feasibility of using a small, flexible, injectable, subcutaneous microelectrode-based device to record electrocardiograms (ECGs). We describe the fabrication process and demonstrate the ease of insertion of the injectable ECG device in vivo swine model. We also demonstrate our device's high-density channel microelectrode array's ability to detect the P, R, and T waves. The amplitude of these waves showed excellent correlation with distance of the bipolar electrodes used to detect them. Given the success of our initial studies, this device has the potential to improve the safety profile of implantable cardiac monitors and simplify the implantation procedure to allow for placement in a primary care setting.