Over 27 million individuals are affected every year worldwide with central nervous system (CNS) injuries. These injuries include but are not limited to traumatic brain injury (TBI) and spinal cord injury (SCI). CNS injuries remain a significant public health concern which demands reliable tools for rapid, on-sight, on-field, and point-of-care diagnostic (POC) solutions. To address these challenges, we developed a low-cost, open-source, hand-held, portable, and POC detection technology, termed as MicroDrop (μDrop), which can simultaneously detect up to eight target biomolecules and display results in both analog and digital formats. The data acquired is stored wirelessly in a cloud server for further investigation and statistical analysis. Multiplexing capability of μDrop and immuno-biosensors detects and quantifies Cleaved-Tau Protein (C-Tau) and Neuron-Filament (NFL) proteins in the blood of TBI patients. Immuno-biosensors rapidly sense the two target proteins in less than 30 min, with μDrop and a conventional potentiostat. C-Tau and NFL were selectively detected with μDrop within the dynamic range of 10 pg/mL - 100 ng/mL and the sensitivity range of 47 μA/pg mm2 - 65 μA/pg mm2. Comparing the biosensing performance with enzyme-linked immunosorbent assays (ELISA) shows that the immuno-biosensors combined with μDrop could successfully differentiate between clinical controls and injured patients.