Evaluation of three rapid lateral flow antigen detection tests for the diagnosis of SARS-CoV-2 infection.

Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology Jääskeläinen AE, Ahava MJ, Jokela P, Szirovicza L, Pohjala S, Vapalahti O, Lappalainen M, Hepojoki J, Kurkela S
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

INTRODUCTION: The COVID-19 pandemic has led to high demand of diagnostic tools. Rapid antigen detection tests have been developed and many have received regulatory acceptance such as CE IVD or FDA markings. Their performance needs to be carefully assessed. MATERIALS AND METHODS: 158 positive and 40 negative retrospective samples collected in saline and analyzed by a laboratory-developed RT-PCR test were used to evaluate Sofia (Quidel), Standard Q (SD Biosensor), and Panbio™ (Abbott) rapid antigen detection tests (RADTs). A subset of the specimens was subjected to virus culture. RESULTS: The specificity of all RADTs was 100 % and the sensitivity and percent agreement was 80 % and 85 % for Sofia, 81 % and 85 % for Standard Q, and 83 % and 86 % for Panbio™, respectively. All three RADTs evaluated in this study reached a more than 90 % sensitivity for samples with a high viral load as estimated from the low Ct (Cycle threshold) values in the reference RT-PCR. Virus culture was successful in 80 % of specimens with a Ct value <25. CONCLUSIONS: As expected, the RADTs were less sensitive than RT-PCR. However, they benefit from the speed and ease of testing, and lower price as compared to RT-PCR. Repeated testing in appropriate settings may improve the overall performance.