Creatinine is an indicator of hindrance in urination and renal insufficiency. Creatinine levels are the marker of the late stages of prostate cancer. Early and sensitive detection of creatinine can reduce deaths associated with prostate cancer. In this work, nitrogen-doped porous carbon antimony (Sb/NPC) nanoparticles are fabricated to be employed as a non-enzymatic biosensor. Sb/NPC has promising redox activity and is synthesized by a two-step reaction using low-cost precursors. Electrochemical sensing by Sb/NPC is conducted for standard creatinine solutions on a three-electrodes system. Cyclic voltammetry, amperometry, and electrochemical impedance spectroscopy are used to sense creatinine. LOD and LOQ of the Sb/NPC modified electrode are 0.74 µM and 2.4 µM, respectively. This electrode system analyzes creatinine in the serum of prostate cancer patients who have elevated PSA levels. More than 90% creatinine is recovered from a spiked serum sample of a prostate cancer patient. A direct relation is observed between PSA levels and creatinine levels in prostate cancer. The developed cyclic voltammetric setup detects trace concentrations of creatinine in serum.