Anti-pesticide DNA aptamers fail to recognize their targets with asserted micromolar dissociation constants.

Analytica chimica acta Zara L, Achilli S, Chovelon B, Fiore E, Toulmé JJ, Peyrin E, Ravelet C
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Herein, we originally aimed at developing fluorescence anisotropy biosensor platforms devoted to the homogeneous-phase detection of isocarbophos and phorate pesticides by using previously isolated DNA aptamers. To achieve this, two reporting approaches displaying very high generalizability features were implemented, based on either the complementary strand or the SYBR green intercalator displacement strategies. Unfortunately, none of the transduction methods led to phorate-dependent signals. Only the SYBR green displacement method provided a small output in the presence of isocarbophos, but at an analyte concentration greater than 100 μM. In order to identify the origin of such data, isothermal titration calorimetry (ITC) experiments were subsequently performed. It was shown that aptamers bind neither isocarbophos nor phorate in free solution with the claimed micromolar dissociation constants. This work puts forward some doubts about the previously described aptasensors that rely on the use of these functional DNA molecules. It also highlights the need to carefully investigate the binding capabilities of aptamers after their isolation and to include appropriate control experiments with scrambled or mutated oligonucleotides.