2021

Antifouling Peptide Hydrogel Based Electrochemical Biosensors for Highly Sensitive Detection of Cancer Biomarker HER2 in Human Serum.

Analytical chemistry Wang W, Han R, Chen M, Luo X
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A facile strategy for the electrochemical detection of human epidermal growth factor receptor 2 (HER2), a breast cancer biomarker, was presented via the fabrication of an antifouling sensing interface based on the conducting polymer poly(3,4-ethylenedioxythiophene) (PEDOT) and a biocompatible peptide hydrogel. The peptide hydrogel prepared from a designed short peptide of Phe-Glu-Lys-Phe functionalized with a fluorene methoxycarbonyl group (Fmoc-FEKF) enabled excellent activity preservation for the immobilized biomolecules, and its good hydrophilicity facilitated effective alleviation of nonspecific adsorption or biofouling, while the PEDOT film provided a highly stable and conducting substrate. The developed biosensor was highly sensitive and selective for HER2 detection, with a wide linear response range from 0.1 ng mL-1 to 1.0 μg mL-1 and a low limit of detection of 45 pg mL-1. Moreover, the peptide hydrogel based biosensor was feasible to use for complex biological samples, and it was capable of detecting HER2 in human serum with clinically acceptable accuracy, manifesting a promising potential for practical application.

关键词