Zwitterions for impedance spectroscopy: The new buffers in town.

Analytica chimica acta Anand S, Swami P, Goel G, Gupta S
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Studying the role of buffers in impedance spectroscopy is a relatively unexplored area. We demonstrate a special class of biologically relevant buffers known as Good's zwitterionic buffers that show improved performance over standard electrolyte buffers (e.g. PBS) currently widely used in impedance spectroscopy measurements of bacterial suspensions. Our theoretical and experimental comparisons of conductivity of classical and zwitterionic buffers at various different concentrations show that ion-ion interaction effects are significantly higher in zwitterionic buffers as compared to classical buffers at the concentrations at which they are used. This and the fact that zwitterions have larger sizes leads to the lowering of their conductivity which significantly improves their impedance sensing ability. We illustrate through an example of heat-induced ionic release in model S. typhi and S. aureus bacteria that having a low conductivity buffer is indeed beneficial for biological impedance measurements. In fact, the best buffer for impedance studies can be chosen solely based on their electrical properties as long as they are also biologically compatible. This gives Good's zwitterionic buffers an edge over conventional media as they satisfy both these criteria.

关键词