2021

Nanoporous Gold for the Miniaturization of In Vivo Electrochemical Aptamer-Based Sensors.

ACS sensors Downs AM, Gerson J, Hossain MN, Ploense K, Pham M, Kraatz HB, Kippin T, Plaxco KW
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Electrochemical aptamer-based sensors enable real-time molecular measurements in the living body. The spatial resolution of these measurements and ability to perform measurements in targeted locations, however, is limited by the length and width of the device's working electrode. Historically, achieving good signal to noise in the complex, noisy in vivo environment has required working electrode lengths of 3-6 mm. To enable sensor miniaturization, here we have enhanced the signaling current obtained for a sensor of given macroscopic dimensions by increasing its surface area. Specifically, we produced nanoporous gold via an electrochemical alloying/dealloying technique to increase the microscopic surface area of our working electrodes by up to 100-fold. Using this approach, we have miniaturized in vivo electrochemical aptamer-based (EAB) sensors (here using sensors against the antibiotic, vancomycin) by a factor of 6 while retaining sensor signal and response times. Conveniently, the fabrication of nanoporous gold is simple, parallelizable, and compatible with both two- and three-dimensional electrode architectures, suggesting that it may be of value to a range of electrochemical biosensor applications.