Free Myocutaneous Flap Assessment in a Rat Model: Verification of a Wireless Bioelectrical Impedance Assessment (BIA) System for Vascular Compromise Following Microsurgery.

Journal of personalized medicine Huang YK, Wong MY, Wu CR, Cheng YZ, Lin BS
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

BACKGROUND: Microvascular tissue transfer is a common reconstructive procedure. We designed a bioelectrical impedance assessment (BIA) system for quantitative analysis of tissue status. This study attempts to verify it through the animal model. METHODS: The flaps of the rat model were monitored by the BIA system. RESULTS: The BIA variation of the free flap in the rat after the vascular compromise was recorded. The non-vascular ligation limbs of the same rat served as a control group. The bio-impedance in the experimental group was larger than the control group. The bio-impedances of both the thigh/feet flaps in the experimental group were increased over time. In the thigh, the difference in bio-impedance from the control group was first detected at 10 kHz at the 3rd and last at 1 kHz at the 6th h, after vascular compromise. The same finding was observed in the feet. Compared with the control group, the bio-impedance ratio (1 kHz/20 kHz) of the experimental group decreased with time, while their variation tendencies in the thigh and feet were similar. CONCLUSIONS: The flap may be monitored by the BIA for vascular status.