2021

A spore-based portable kit for on-site detection of fluoride ions.

Journal of hazardous materials Qin Y, Faheem A, Hu Y
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The excess residues of fluoride ions cause serious human health problems, making their detection highly valuable. In this work, a whole-cell-based biosensor was presented for the detection of fluoride ions, which can inhibit the color reaction of 3,3',5,5',-tetramethylbenzidine (TMB) catalyzed by the CotA-laccase of spore surface. This reaction for the detection of fluoride ions could be read out through UV-vis spectrophotometer, smartphone, or standard colorimetric card within 10 min. Under optimum conditions, a linear range of 1-600 μmol L-1 with a detection limit of 0.12 μmol L-1 (3σ/k) was achieved for fluoride ions detection by using UV-vis spectrophotometer. The biosensor coupling with smartphone had a good linear response to fluoride ions concentration in the range of 5-600 μmol L-1 with LOD of 0.90 μmol L-1 (3σ/k). The standard colorimetric card can be directly used for recognizing the fluoride ions level via naked-eyes. A portable kit based on a colorimetric card and smartphone was developed and has been successfully applied for fluoride ions monitoring in surface waters and groundwater. This developed method has several advantages such as rapid, outstanding selectivity and anti-interference, low-cost, ease of operation and storage, and eco-friendliness, meeting the demands of point-of-care testing of fluoride ions and disease prevention.

关键词