2021

Measurement of Low Concentration of Micro-Plastics by Detection of Bioaffinity-Induced Particle Retention Using Surface Plasmon Resonance Biosensors.

Biosensors Huang CJ, Narasimha GV, Chen YC, Chen JK, Dong GC
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The issue of micro-plastics is becoming more and more important due to their ubiquity and the harm they cause to the human body. Therefore, evaluating the biological-physical interaction of micro-plastics with health cells has become the focus of many research efforts. This study focuses on the movement mode and low concentration detection development for micro-plastics in surface plasmon resonance (SPR). Firstly, 20-micrometer micro-plastics were prepared by grinding and filtering, and the movement mode was explored; then, the characteristics were investigated by SPR. Chromatographic analysis showed that the surface charge of micro-plastics dominated the elution time, and estrogen receptors (ERs) played a supporting role. A difference of micro-plastics in SPR sensorgram was observed, inferring the micro-plastics' movement in rolling mode on the ERs. Characteristics analysis indicated that the low particle number of micro-plastics on SPR showed a linear relationship with the response unit (RU). When ERs were immobilized on the biosensor, the force of the binding of micro-plastics to ERs under an ultra-low background was equivalent to the dissociation rate constant shown as follows: PS (0.05 nM) > PVC (0.09 nM) > PE (0.14 nM). The ELISA-like magnetic beads experiment verified the specificity between ERs and micro-plastics. Therefore, by using the SPR technique, a biological-derived over-occupation of PS was found via higher binding force with ERs and longer retention time. In the future, there will be considerable potential for micro-plastics issues, such as identification in natural samples, biomarking, real-time detection in specific environments/regions and human health subject.

关键词