2022

Design and Performance Analysis of Reconfigurable 1D Photonic Crystal Biosensor Employing Ge₂Sb₂Te₅ (GST) for Detection of Women Reproductive Hormones.

IEEE transactions on nanobioscience Panda A, Vigneswaran D, Pukhrambam PD, Ayyanar N, Nguyen TK
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The present research demonstrates a novel 1D photonic crystal (PhC) based reconfigurable biosensor pertaining to label-free detection of different concentrations of progesterone and estradiol, which play a vital role in developing reproductive hormones in women. The proposed sensor is designed by an alternative arrangement of Na3 AlF6 and CeO2, with a central defect layer. A thin layer of novel phase change chalcogenide material (Ge2 Sb2 Te5) is deposited along the two sides of the defect layer to improve the sensing performance. Numerical simulation of transmission spectrum for TE mode is carried out by using the transfer matrix method (TMM). The mainstay of this research is centered on the assay of shift in the defect mode position and intensity with respect to different concentrations of analyte, by changing the phase of the GST material from amorphous to crystalline. Interestingly, we observed a high tunability in defect mode wavelength, when the phase is changed from amorphous to crystalline, which leads to accomplishment of a high sensitivity of 1.75 nm/nmol/L for progesterone and 20.5 nm/nmol/L for estradiol. Aside from sensitivity, other significant parameters like figure of merit and detection limit are computed, which give a deep insight into the sensing performance. These encouraging sensing performances pave the path for efficient detection of different concentrations of progesterone and estradiol to monitor various gynecological problems in women.

关键词