2021

Parallel Potentiometric and Capacitive Response in a Water-Gate Thin Film Transistor Biosensor at High Ionic Strength.

Sensors (Basel, Switzerland) AlQahtani H, Alswieleh A, Al-Khurayyif I, AlGarni S, Grell M
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

We show that an SnO2-based water-gate thin film transistor (WGTFT) biosensor responds to a waterborne analyte, the spike protein of the SARS-CoV-2 virus, by a parallel potentiometric and capacitive mechanism. We draw our conclusion from an analysis of transistor output characteristics, which avoids the known ambiguities of the common analysis based on transfer characteristics. Our findings contrast with reports on organic WGTFT biosensors claiming a purely capacitive response due to screening effects in high ionic strength electrolytes, but are consistent with prior work that clearly shows a potentiometric response even in strong electrolytes. We provide a detailed critique of prior WGTFT analysis and screening reasoning. Empirically, both potentiometric and capacitive responses can be modelled quantitatively by a Langmuir‒Freundlich (LF) law, which is mathematically equivalent to the Hill equation that is frequently used for biosensor response characteristics. However, potentiometric and capacitive model parameters disagree. Instead, the potentiometric response follows the Nikolsky-Eisenman law, treating the analyte 'RBD spike protein' as an ion carrying two elementary charges. These insights are uniquely possible thanks to the parallel presence of two response mechanisms, as well as their reliable delineation, as presented here.

关键词