2021

Smartphone-Based Chemiluminescent Origami µPAD for the Rapid Assessment of Glucose Blood Levels.

Biosensors Calabria D, Zangheri M, Trozzi I, Lazzarini E, Pace A, Mirasoli M, Guardigli M
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Microfluidic paper analytical devices (µPADs) represent one of the most appealing trends in the development of simple and inexpensive analytical systems for diagnostic applications at the point of care (POC). Herein, we describe a smartphone-based origami µPAD for the quantitative determination of glucose in blood samples based on the glucose oxidase-catalyzed oxidation of glucose leading to hydrogen peroxide, which is then detected by means of the luminol/hexacyanoferrate(III) chemiluminescent (CL) system. By exploiting the foldable µPAD format, a two-step analytical procedure has been implemented. First, the diluted blood sample was added, and hydrogen peroxide was accumulated, then the biosensor was folded, and a transport buffer was added to bring hydrogen peroxide in contact with CL reagents, thus promoting the CL reaction. To enable POC applicability, the reagents required for the assay were preloaded in the µPAD so that no chemicals handling was required, and a 3D-printed portable device was developed for measuring the CL emission using the smartphone's CMOS camera. The µPAD was stable for 30-day storage at room temperature and the assay, displaying a limit of detection of 10 µmol L-1, proved able to identify both hypoglycemic and hyperglycemic blood samples in less than 20 min.