2021

An impedimetric approach for COVID-19 detection.

The Analyst Tepeli Büyüksünetçi Y, Çitil BE, Anık Ü
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

In this study, an electrochemical approach for the determination of coronavirus disease (COVID-19) was developed. The biosensor system relied on the spike protein (S-protein) based infection mechanism of the virus and included separate interactions of receptors like angiotensin-converting enzyme 2 (ACE2) and CD147. After the optimization of experimental parameters, the analytical characteristics of both receptors ACE2 and CD147 were investigated. For ACE2 receptor, the linear detection ranges of the S-protein were found in the range of 700 ng mL-1 to 1500 ng mL-1 and from 1500 ng mL-1 to 7000 ng mL-1 with a limit of detection (LOD) value of 299.30 ng mL-1. Meanwhile, for CD147 receptor the linear range was in the range of 500 ng mL-1 to 5000 ng mL-1 with a LOD value of 38.99 ng mL-1. After the examination of analytical characteristics, the developed electrochemical approach was applied for severe acute respiratory syndrome coronavirus 2 samples and the obtained results were validated with real time polymerase chain reaction method.

关键词