2022

Continual and accurate home monitoring of uric acid in urine samples with uricase-packaged nanoflowers assisted portable electrochemical uricometer.

Biosensors & bioelectronics Xing X, Yao B, Wu Q, Zhang R, Yao L, Xu J, Gao G, Chen W
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A convenient, fast and non-invasive portable electrochemical uricometer (PUM) assisted with the uricase-packaged nanoflowers (NFs) was constructed for continually and accurately monitoring of uric acid (UA) in urine samples at random intervals in just 20 s. Only a small amount of urine (50 μL) was needed for each test. Electrochemical deposition was adopted to modify gold nanoparticles (AuNPs) on screen-printed carbon electrodes (SPCE) and uricase-inorganic hybrid NFs (UOx-NFs) induced by calcium ions (Ca2+) were introduced for UA detection with expected specificity. Cyclic voltammetry (CV) (detection limit of 8.87 μM and liner range of 0-4 mM) and amperometry (detection limit of 0.82 μM and liner range of 0-5 mM) protocols were studied for UA detection, respectively. Finally, the uric acid in urine had be successfully continually monitored from volunteers with various dietary choosing, the results of which can be adopted as the effective evidence for uric acid control.

关键词