2021

Nanowired dual-electrodes surface to monitor cerebral ischemia by current-volt measurements.

3 Biotech Gong W, Sun M, Guo X, Liu Y, Li H, Xie L, Li X
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The level of clotting protein 'factor IX' (FIX) is highly associated with cerebral ischemia, and this research work has developed a sensitive detection of FIX on dielectrode sensor by current-volt measurement. Sensing area was grown with zinc oxide nanowire to attach more probe for FIX interaction. Aptamer was utilized as the detection probe and attached on the sensing electrode surface through amine-aldehyde chemical linkage. In addition, biotin-streptavidin interaction was utilized to attach the higher number aptamers on the electrode surface connected with dual-probe station. FIX detection limit was found as 10 fM in the phosphate buffer saline spiked samples and 1:320 dilution of human serum. The linear ranges were as 10 fM to 100 pM and 1:320 to 1:80, respectively. With a good determination co-efficient [y = 2.6813x - 3.8467; R 2 = 0.9479] this biosensing strategy helps to quantify FIX and monitor the condition of cerebral ischemia.

关键词