Point-of-care testing plays an important role in the detection of skin abnormalities. The detection of skin abnormalities requires sufficient depth and no harm. A flexible microwave biosensor based on spoof surface plasmon polaritons was designed to meet the requirements of skin abnormalities. The designed biosensor, which works at 11.3 GHz, is small and can be flexibly attached to the skin surface of any part of the human body for measurement. The health status of the skin can be evaluated by the resonant frequency and the magnitude of the reflection coefficient of the sensor. The sensor was tested on pork skin. The experiment results showed that the sensor can detect skin abnormalities such as skin burn, skin tumor, and others. Compared with other sensors, the sensor has sufficient penetration depth because of the strong penetration of microwave electromagnetic waves. It is the first flexible microwave biosensor used for skin, which involves point-of-care testing, and continuous monitoring of skin.