2022

Glutamate Oxidase-Integrated Biomimetic Metal-Organic Framework Hybrids as Cascade Nanozymes for Ultrasensitive Glutamate Detection.

Journal of agricultural and food chemistry Hu H, Li P, Wang Z, Du Y, Kuang G, Feng Y, Jia S, Cui J
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The hybrid coupling of biocatalysts and chemical catalysts plays a vital role in the fields of catalysis, sensing, and medical treatment due to the integrated advantages in the high activity of natural enzymes and the excellent stability of nanozymes. Herein, a new nanozyme/natural enzyme hybrid biosensor was established for ultrasensitive glutamate detection. The MIL-88B(Fe)-NH2 material with remarkable peroxidase mimic activity and stability was used as a nanozyme and carrier for immobilizing glutamate oxidase (GLOX) through Schiff base reaction to construct a chem-enzyme cascade detector (MIL-88B(Fe)-NH2@GLOX). The resultant MIL-88B(Fe)-NH2@GLOX exhibited a wide linear range (1-100 μM), with a low detection limit of 2.5 μM for glutamate detection. Furthermore, the MIL-88B(Fe)-NH2@GLOX displayed excellent reusability and storage stability. After repeated seven cycles, MIL-88B(Fe)-NH2-GLOX (GLOX was adsorbed on MIL-88B(Fe)-NH2) lost most of its activity, whereas MIL-88B(Fe)-NH2@GLOX still retained 69% of its initial activity. Meanwhile, MIL-88B(Fe)-NH2@GLOX maintained 60% of its initial activity after storage for 90 days, while free GLOX only retained 30% of its initial activity. This strategy of integrating MOF mimics and natural enzymes for cascade catalysis makes it possible to design an efficient and stable chemo-enzyme composite catalysts, which are promising for applications in biosensing and biomimetic catalysis.

关键词