2022

Closed-Loop Diabetes Minipatch Based on a Biosensor and an Electroosmotic Pump on Hollow Biodegradable Microneedles.

ACS sensors Luo X, Yu Q, Liu Y, Gai W, Ye L, Yang L, Cui Y
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Developing a miniaturized, low-cost, and smart closed-loop system for diabetes could significantly improve life quality and benefit millions of people. Conventional closed-loop devices are large in size and exorbitant. Here, we unprecedentedly demonstrate an electrically controlled flexible closed-loop patch for continuous diabetes management by integrating hollow biodegradable microneedles with a biosensing device and an electroosmotic pump. The hollow microneedles were fabricated using a combination of soft lithography and micromachining. The outer layer of the microneedles was functionalized to serve as a biosensing device for the in situ sensitive and accurate monitoring of interstitial glucose. The inner layer of the microneedles was integrated with a flexible electroosmotic pump to deliver insulin, and the delivery rate was electrically controlled by the glucose level from the biosensing device. The closed-loop system successfully stabilized the blood glucose levels of diabetic rats in a normal and safe range. The system is painless, miniaturized, cost-effective, and flexible. It is anticipated that it could open up exciting new avenues for fundamental studies of new closed-loop devices as well as practical applications for diabetes management.

关键词