2022

Electronic field effect detection of SARS-CoV-2 N-protein before the onset of symptoms.

Biosensors & bioelectronics Novodchuk I, Kayaharman M, Prassas I, Soosaipillai A, Karimi R, Goldthorpe IA, Abdel-Rahman E, Sanderson J, Diamandis EP, Bajcsy M, Yavuz M
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

As part of the efforts to contain the pandemic, researchers around the world have raced to develop testing platforms to detect the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), which causes the Coronavirus disease 2019 (COVID-19). Within the different detection platforms studied, the field effect transistor (FET) is a promising device due to its high sensitivity and fast detection capabilities. In this work, a graphene-based FET which uses a boron and nitrogen co-doped graphene oxide gel (BN-GO gel) transducer functionalized with nucleoprotein antibodies, has been investigated for the detection of SARS-CoV-2 nucleocapsid (N)-protein in buffer. This biosensor was able to detect the viral protein in less than 4 min, with a limit of detection (LOD) as low as 10 ag/mL and a wide linear detection range stretching over 11 orders of magnitude from 10 ag/mL-1 μg/mL. This represents the lowest LOD and widest detection range of any COVID-19 sensor and thus can potentially enable the detection of infected individuals before they become contagious. In addition to its potential use in the COVID-19 pandemic, our device serves as a proof-of-concept of the ability of functionalized BN-GO gel FETs to be used for ultrasensitive yet robust biosensors.

关键词