2022

d-Amino Acid-Based Antifouling Peptides for the Construction of Electrochemical Biosensors Capable of Assaying Proteins in Serum with Enhanced Stability.

ACS sensors Zhao S, Qiao X, Chen M, Li Y, Wang X, Xu Z, Wu Y, Luo X
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The susceptibility of peptides to proteolytic degradation in human serum significantly hindered the potential application of peptide-based antifouling biosensors for long-term assaying of clinical samples. Herein, a robust antifouling biosensor with enhanced stability was constructed based on peptides composed of d-amino acids (d-peptide) with prominent proteolytic resistance. The electrode was electropolymerized with poly(3,4-ehtylenedioxythiophene) and electrodeposited with Au nanoparticles (AuNPs), and the d-peptide was then immobilized onto the AuNPs, and a typical antibody specific for immunoglobulin M (IgM) was immobilized. Because of the effect of d-amino acids, the d-peptide-modified electrode surface showed prominent antifouling capability and high tolerance to enzymatic hydrolysis. Moreover, the d-peptide-modified electrode exhibited much stronger long-term stability, as well as antifouling ability in human serum than the electrode modified with normal peptides. The electrochemical biosensor exhibited a sensitive response to IgM linearly within the range of 100 pg mL-1 to 1.0 μg mL-1 and a very low detection limit down to 37 pg mL-1, and it was able to detect IgM in human serum with good accuracy. This work provided a new strategy to develop robust peptide-based biosensors to resist the proteolytic degradation for practical application in complex clinical samples.

关键词