Wearables are valuable solutions for monitoring a variety of physiological parameters. Their application in cardiorespiratory monitoring may significantly impact global health problems and the economic burden related to cardiovascular and respiratory diseases. Here, we describe a soft biosensor capable of monitoring heart (HR) and respiratory (RR) rates simultaneously. We show that a skin-interfaced biosensor based on fiber optics (i.e., the smart patch) is capable of estimating HR and RR by detecting local ribcage strain caused by breathing and heart beating. The system addresses some of the main technical challenges that limit the wide-scale use of wearables, such as the simultaneous monitoring of HR and RR via single sensing modalities, their limited skin compliance, and low sensitivity. We demonstrate that the smart patch estimates HR and RR with high fidelity under different respiratory conditions and common daily body positions. We highlight the system potentiality of real-time cardiorespiratory monitoring in a broad range of home settings.