2022

Development of a Bacterial Enzyme-Based Biosensor for the Detection and Quantification of Selenate.

Chemistry (Weinheim an der Bergstrasse, Germany) Motlagh MK, Noroozifar M, Sodhi RNS, Kraatz HB
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

An enzymatic biosensor has been developed for the determination of selenate (SeO4 2- ), in which selenate reductase (SeR) is chemically attached to a gold disk electrode by lipoic acid N-hydroxysuccinimide ester as linker, allowing the catalytic reduction of the SeO4 2- to SeO3 2- . Modification of the gold electrode was characterized by X-ray photoelectron spectroscopy (XPS), time-of-flight secondary ion mass spectroscopy (ToF-SIMS), and electrochemistry. Cyclic voltammetry (CV) and differential pulse voltammetry (DPV) measurements were performed in different buffers for selenate determination. Under optimum conditions, the calibration curve was linear over the range 7.0-3900.0 μg L-1 with limits of detection and quantification of 4.97 and 15.56 μg L-1 , respectively. The possible interference of the relevant oxyanions SO4 2- , NO3 - , NO2 - , PO4 3- and AsO4 3- in the determination of SeO4 2- was studied. Finally, the proposed biosensor was used to determine SeO4 2- with recovery between 95.2 and 102.4 % in different real water samples.