2022

Electrochemical DNA biosensor for chronic myelocytic leukemia based on hybrid nanostructure.

Bioelectrochemistry (Amsterdam, Netherlands) Avelino KYPS, Oliveira LS, Santos MR, Lucena-Silva N, Andrade CAS, Oliveira MDL
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The present research refers to elaborating a new label-free electrochemical biosensor used to detect the BCR/ABL fusion gene. We used a hybrid nanocomposite composed of chitosan and zinc oxide nanoparticles (Chit-ZnONP) immobilized on a polypyrrole (PPy) film. DNA segments were covalently immobilized, allowing biomolecular recognition. Atomic force microscopy (AFM), cyclic voltammetry (CV), and electrochemical impedance spectroscopy (EIS) were used to evaluate the assembly stages of the biosensor. The biosensor's analytical performance was investigated using recombinant plasmids containing the target oncogene and clinical samples from patients with chronic myeloid leukemia (CML). A limit of detection (LOD) of 1.34 fM, limit of quantification (LOQ) of 4.08 fM, and sensitivity of 34.03 μA fM-1 cm2 were calculated for the BCR/ABL fusion oncogene. The sensing system exhibited high specificity, selectivity, and reproducibility with a standard deviation (SD) of 4.21%. Additionally, a linear response range was observed between 138.80 aM to 13.88 pM with a regression coefficient of 0.96. Also, the biosensor shows easy operationalization and fast analytical response, contributing to the early cancer diagnosis. The proposed nanostructured device is an alternative for the genetic identification BCR/ABL fusion gene.

关键词