2022

Enzyme sensor for simultaneous determination of cholesterol, triglyceride and HDL cholesterol in serum.

Analytical sciences : the international journal of the Japan Society for Analytical Chemistry Uemura T, Yamazaki H, Itoh T, Nishizawa S
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A three-electrode lipid biosensor that simultaneously measures the total cholesterol (CHO), triglyceride (TG), and HDL cholesterol (HDL-C) in standard serum has been developed. The lipid biosensor is designed for clinical use where production cost, low sample requirement, portability, stability, and speed are high priorities. The device design filters out blood cells and lipoproteins from the serum, where the target molecules are catalyzed by enzymes encapsulated in carboxymethyl cellulose (CMC) to produce electron mediators, potassium ferrocyanide. These electron mediators were subsequently detected by amperometric determination. The sensor exhibit high selectivity towards the targets and can measure the target lipids in 4 min with 10 µL of serum. Lastly, the device can be stored up to 18 months with a minimal decrease in catalytic efficiency.

关键词