2022

Aptasensor for Detection of Influenza-A in Human Saliva.

Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference Maddocks GM, Peterson KL, Downey ML, Lee BH, Lavoie JH, Menegatti S, Daniele M
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Access to low-cost, rapid, individualized diagnostics at point-of-care and point-of-need is vital to minimize the impact of highly infectious viruses, such as influenza. Herein, a biosensor for detecting hemagglutinin (HA), an abundant capsid protein in H1N1 viruses, is demonstrated. A gold working electrode was functionalized with a thiol-modified, HA-binding aptamer derivatized with a methylene blue modification for redox reporting. The aptamer was characterized by surface plasmon resonance to confirm its biorecognition activity for HA. The aptasensor was characterized by square wave voltammetry to quantify the sensor's response to varying concentrations of HA. The sensor exhibited a lower limit of detection of 1.5 pM with linear detection of up to 1.2 nM in both Tris buffer and simulated human saliva, thus encompassing the clinically relevant HA range in saliva. Average sensitivity was measured at 21.083 nA·nM-1in Tris and 14.5 nA·nM-1in artificial saliva across clinically relevant HA titers. Sensor stability across time was also investigated, providing a preliminary understanding of the translational viability of the aptasensors for mobile and remote diagnostic applications.

关键词