2022

Catalytic Biosensors Operating under Quasi-Equilibrium Conditions for Mitigating the Changes in Substrate Diffusion.

Angewandte Chemie (International ed. in English) Muhs A, Bobrowski T, Lielpētere A, Schuhmann W
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Despite the success of continuous glucose measuring systems operating through the skin for about 14 days, long-term implantable biosensors are facing challenges caused by the foreign-body reaction. We present a conceptually new strategy using catalytic enzyme-based biosensors based on a measuring sequence leading to minimum disturbance of the substrate equilibrium concentration by controlling the sensor between "on" and "off" state combined with short potentiometric data acquisition. It is required that the enzyme activity can be completely switched off and no parasitic side reactions allow substrate turnover. This is achieved by using an O2 -independent FAD-dependent glucose dehydrogenase embedded within a crosslinked redox polymer. A short measuring interval allows the glucose concentration equilibrium to be restored quickly which enables the biosensor to operate under quasi-equilibrium conditions.

关键词