组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
In this paper, we present the results of a non-enzymatic electrochemical glucose biosensor based on TiO2. An anatase working electrode was synthesized using the spin coating technique with the polymeric precursor method and dispersed TiO2 nanoparticles. Through scanning electron microscopy, it was observed that the electrode presented an irregular surface with clusters of nanoparticles. Electrochemical characterization indicated that the response was directly related to the morphology of the electrode. In the presence of glucose, the electrode exhibited adsorption behavior toward the molecules, enabling their recognition. The electrode was tested by employing PBS (phosphate buffer solutions) with varying pH values (from 4 to 9), demonstrating its electrochemical stability, even in the presence of glucose. Amperometric characterization was used to determine that the working region appeared from 0.2 mM to 2 mM, with a sensitivity of 4.46 μAcm-2mM-1 in PBS pH 7. The obtained results suggest that TiO2-based electrodes could be used for the detection of glucose concentration in sweat (0.277-1 mM) and saliva (0.23-1.77 mM).