2022

Multiple Biomarker Simultaneous Detection in Serum via a Nanomaterial-Functionalized Biosensor for Ovarian Tumor/Cancer Diagnosis.

Micromachines Wu Y, Wang C, Wang C, Wang P, Chang X, Han L, Zhang Y
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Ovarian tumors/cancers are threatening women's health worldwide, which demands high-performance detection methods and accurate strategies to effectively detect, diagnose and treat them. Here, we report a nanographene oxide particle-functionalized microfluidic fluorescence biosensor to simultaneously detect four biomarkers, CA125, HE4, CEA and APF, for ovarian tumor/cancer diagnosis. The developed biosensor exhibits good selectivity and a large biomarker detection range with a limit of detection of 0.01 U/mL for CA125 and ~1 pg/mL for HE4, CEA and APF. The current results indicate that (1) the proposed biosensor is a promising tool for the simultaneous detection of multiple biomarkers in ovarian tumors/cancer and (2) CA125 and HE4 are strong indicators, AFP may be helpful, and CEA is a weak biomarker for ovarian tumor/cancer diagnosis. The proposed biosensor would be a potential tool, and an analytical approach for the simultaneous detection of multiple biomarkers will provide a new strategy for the early screening, diagnosis and treatment of ovarian tumors/cancers, as well as other cancers.