Head-to-head comparison of nasal and nasopharyngeal sampling using SARS-CoV-2 rapid antigen testing in Lesotho.

PloS one Labhardt ND, González Fernández L, Katende B, Muhairwe J, Bresser M, Amstutz A, Glass TR, Ruhwald M, Sacks JA, Escadafal C, Mareka M, Mooko SM, de Vos M, Reither K
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

OBJECTIVES: To assess the real-world diagnostic performance of nasal and nasopharyngeal swabs for SD Biosensor STANDARD Q COVID-19 Antigen Rapid Diagnostic Test (Ag-RDT). METHODS: Individuals ≥5 years with COVID-19 compatible symptoms or history of exposure to SARS-CoV-2 presenting at hospitals in Lesotho received two nasopharyngeal and one nasal swab. Ag-RDT from nasal and nasopharyngeal swabs were performed as point-of-care on site, the second nasopharyngeal swab used for polymerase chain reaction (PCR) as the reference standard. RESULTS: Out of 2198 participants enrolled, 2131 had a valid PCR result (61% female, median age 41 years, 8% children), 84.5% were symptomatic. Overall PCR positivity rate was 5.8%. The sensitivity for nasopharyngeal, nasal, and combined nasal and nasopharyngeal Ag-RDT result was 70.2% (95%CI: 61.3-78.0), 67.3% (57.3-76.3) and 74.4% (65.5-82.0), respectively. The respective specificity was 97.9% (97.1-98.4), 97.9% (97.2-98.5) and 97.5% (96.7-98.2). For both sampling modalities, sensitivity was higher in participants with symptom duration ≤ 3days versus ≤ 7days. Agreement between nasal and nasopharyngeal Ag-RDT was 99.4%. CONCLUSIONS: The STANDARD Q Ag-RDT showed high specificity. Sensitivity was, however, below the WHO recommended minimum requirement of ≥ 80%. The high agreement between nasal and nasopharyngeal sampling suggests that for Ag-RDT nasal sampling is a good alternative to nasopharyngeal sampling.

关键词