2023

Optical Glucose Sensors Based on Chitosan-Capped ZnS-Doped Mn Nanomaterials.

Sensors (Basel, Switzerland) Nguyen SH, Vu PKT, Nguyen HM, Tran MT
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The primary goal of glucose sensing at the point of care is to identify glucose concentrations within the diabetes range. However, lower glucose levels also pose a severe health risk. In this paper, we propose quick, simple, and reliable glucose sensors based on the absorption and photoluminescence spectra of chitosan-capped ZnS-doped Mn nanomaterials in the range of 0.125 to 0.636 mM glucose corresponding to 2.3 mg/dL to 11.4 mg/dL. The detection limit was 0.125 mM (or 2.3 mg/dL), much lower than the hypoglycemia level of 70 mg/dL (or 3.9 mM). Chitosan-capped ZnS-doped Mn nanomaterials retain their optical properties while improving sensor stability. This study reports for the first time how the sensors' efficacy was affected by chitosan content from 0.75 to 1.5 wt.%. The results showed that 1 %wt chitosan-capped ZnS-doped Mn is the most-sensitive, -selective, and -stable material. We also put the biosensor through its paces with glucose in phosphate-buffered saline. In the same range of 0.125 to 0.636 mM, the sensors-based chitosan-coated ZnS-doped Mn had a better sensitivity than the working water environment.

关键词