Numerical optimization of microfluidic biosensor detection time for the SARS-CoV-2 using the Taguchi method.

Indian journal of physics and proceedings of the Indian Association for the Cultivation of Science (2004) Ben Mariem I, Kaziz S, Belkhiria M, Echouchene F, Belmabrouk H
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The performance of microfluidic biosensor of the SARS-Cov-2 was numerically analyzed through finite element method. The calculation results have been validated with comparison with experimental data reported in the literature. The novelty of this study is the use of the Taguchi method in the optimization analysis, and an L8(25) orthogonal table of five critical parameters-Reynolds number (Re), Damköhler number (Da), relative adsorption capacity (σ), equilibrium dissociation constant (KD), and Schmidt number (Sc), with two levels was designed. ANOVA methods are used to obtain the significance of key parameters. The optimal combination of the key parameters is Re = 10-2, Da = 1000, σ = 0.2, KD = 5, and Sc 104 to achieve the minimum response time (0.15). Among the selected key parameters, the relative adsorption capacity (σ) has the highest contribution (42.17%) to the reduction of the response time, while the Schmidt number (Sc) has the lowest contribution (5.19%). The presented simulation results are useful in designing microfluidic biosensors in order to reduce their response time.