Performance Comparison Between Conventional Fluorescent Spot Test and Quantitative Assay in Detecting G6PD Deficiency in Neonates.

Oman medical journal Halim SA, Bahar R, Abdullah WZ, Zon EM, Yusoff SM
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

OBJECTIVES: Glucose-6-phosphate dehydrogenase (G6PD) deficiency is the most common enzymopathy worldwide. The fluorescent spot test (FST) is the conventional method for screening neonates for G6PD. However, it has limitations and quantitative assays such as the CareStart Biosensor 1 are being increasingly recommended. This study aimed to compare FST and CareStart Bioensor 1 in their ability to detect G6PD levels in neonates. METHODS: This cross-sectional study involved 455 neonates between June and December 2020. Two milliliters of cord blood were analyzed with CareStart Biosensor 1 and dried cord blood spots with FST. Data was recorded and statistically analyzed. Sensitivity, specificity, positive predictive value, and negative predictive value were calculated to determine the performance of FST at specific G6PD cut-off values; Cohen's kappa analysis assessed the agreement between the two methods. RESULTS: The sensitivity of FST at 30% cut-off G6PD activity level was 91.0%, (95% CI: 57.0-100) and specificity of 97.0% (95% CI: 95.0-98.0). At 60% cut-off, the FST sensitivity sharply declined to 29.0% (95% CI: 19.0-40.0) with a specificity of 100% (95% CI: 98.0-100). The overall prevalence of G6PD deficiency was 5.1% as measured by FST and 17.8% by Biosensor 1 (p < 0.001). CONCLUSIONS: In this study, FST missed a significant proportion of cases of intermediate G6PD levels. FST also misclassified several G6PD intermediate individuals as normal, rendering them susceptible to oxidative stress. Biosensor 1 reported a significantly higher prevalence of G6PD deficiency.

关键词