A disposable immunosensor for the detection of salivary MMP-8 as biomarker of periodontitis.

Bioelectrochemistry (Amsterdam, Netherlands) Tortolini C, Gigli V, Angeloni A, Tasca F, Thanh NTK, Antiochia R
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

This work describes the development of a novel voltammetric immunosensor for the detection of salivary MMP-8 at the point-of-care. The electrochemical platform was based on a graphene (GPH) screen-printed electrode (SPE) functionalized by gold-nanospheres (AuNSs) and antibodies against MMP-8 protein (anti-MMP-8). The functionalization with anti-MMP-8 was realized by using 11-mercaptoundecanoic acid (11-MUA), thanks to its ability to give strong sulfur bonds with its -SH end, and to cross-link the -NH2 groups of the antibody molecule with the other -COOH end, using the traditional EDC-NHS method. The voltammetric sensor showed good performances with a linear range of 2.5-300 ng mL-1, a LOD value of 1.0 ± 0.1 ng mL-1 and a sensitivity of 0.05 µA mL cm-2 ng-1. Moreover, the proposed immunosensor was tested in real saliva samples, showing comparable results to those obtained with the conventional ELISA method. The biosensor was single-use and cost-effective and required a small quantity of test medium and a short preparation time, representing a very attractive biosensor for MMP-8 detection in human saliva.