Monitoring levodopa oxidation and reduction reactions using surface plasmon resonance on a nanohole array electrode.

Discover nano Peng HF, Chang CK, Gupta R, Huang JJ
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The traditional method of monitoring the oxidation and reduction of biomedical materials usually relies on electrochemical (EC) measurement techniques. Here, we demonstrate a surface plasmon resonance (SPR) method to monitor the oxidation process. Using levodopa L-dopa as the target analyte, a nanohole sensing plate is embedded in the EC electrode to enhance the oxidation signal and generate SPR. Cyclic voltammetry (CV) measurement was first conducted to understand the baseline of EC response of L-Dopa. Then, the redox reactions were simultaneously monitored through SPR measurements during the CV voltage scan. The results showed that the limit of detection using traditional CV reached 1.47 μM while using EC-SPR, the limit of detection improved to 1.23 μM. Most importantly, we found a strong correlation between CV current profiles and the SPR reflection spectra. Our results facilitate detecting electrochemical reactions using an optical probing method.