2024

Detection of Blood Clots Using a Whole Stent as an Active Implantable Biosensor.

Advanced science (Weinheim, Baden-Wurttemberg, Germany) Kirimi MT, Hoare D, Holsgrove M, Czyzewski J, Mirzai N, Mercer JR, Neale SL
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Many cardiovascular problems stem from blockages that form within the vasculature and often treatment includes fitting a stent through percutaneous coronary intervention. This offers a minimally invasive therapy but re-occlusion through restenosis or thrombosis formation often occurs post-deployment. Research is ongoing into the creation of smart stents that can detect the occurrence of further problems. In this study, it is shown that selectively metalizing a non-conductive stent can create a set of electrodes that are capable of detecting a build-up of material around the stent. The associated increase in electrical impedance across the electrodes is measured, testing the stent with blood clot to mimic thrombosis. It is shown that the device is capable of sensing different amounts of occlusion. The stent can reproducibly sense the presence of clot showing a 16% +/-3% increase in impedance which is sufficient to reliably detect the clot when surrounded by explanted aorta (one sample t-test, p = 0.009, n = 9). It is demonstrated that this approach can be extended beyond the 3D printed prototypes by showing that it can be applied to a commercially available stent and it is believed that it can be further utilized by other types of medical implants.

关键词