Real-world evaluation of a QCM-based biosensor for exhaled air.

Analytical and bioanalytical chemistry Juste-Dolz A, Teixeira W, Pallás-Tamarit Y, Carballido-Fernández M, Carrascosa J, Morán-Porcar Á, Redón-Badenas MÁ, Pla-Roses MG, Tirado-Balaguer MD, Remolar-Quintana MJ, Ortiz-Carrera J, Ibañez-Echevarría E, Maquieira A, Giménez-Romero D
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The biosensor, named "virusmeter" in this study, integrates quartz crystal microbalance technology with an immune-functionalized chip to distinguish between symptomatic patients with respiratory diseases and healthy individuals by analyzing exhaled air samples. Renowned for its compact design, rapidity, and noninvasive nature, this device yields results within a 5-min timeframe. Evaluated under controlled conditions with 54 hospitalized symptomatic COVID-19 patients and 128 control subjects, the biosensor demonstrated good overall sensitivity (98.15%, 95% CI 90.1-100.0) and specificity (96.87%, 95% CI 92.2-99.1). This proof-of-concept presents an innovative approach with significant potential for leveraging piezoelectric sensors to diagnose respiratory diseases.