2024

Highly Efficient Photoelectrochemical Detection of Cystatin C Based on a Core-Shell MOF Nanocomposite with Biomimetic-Catalysis Amplification.

ACS omega Xia M, Yang P, Zhu C, Hu Y, Fang L, Zheng J, Wang X, Li Y
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Cystatin C (CysC) has been proven to be used to diagnose acute kidney injury (AKI) rapidly and sensitively early. Therefore, it is urgent to develop a sensitive, novel, and rapid method for detecting CysC. In this work, a novel photoelectrochemical (PEC) biosensor was designed for ultrasensitive CysC detection. Ti-MOF@DM-LZU1@Au as a photosensitive material was first modified on the ITO electrode surface. Then, Ab1 and CysC were assembled on the electrode via the specific immunoresponse of an antigen and antibody. Lastly, the conjugate Ab2/l-Cys bilayer/l-Cys-hemin/G-quadruplex with self-catalytic enzyme performance, as a signal amplification approach, could further react with CysC and Ab1, which resulted in a stronger photocurrent. As expected, the constructed PEC sensor realized the ultrasensitive detection of CysC, with a detection range of 10 pg/mL to 16 μg/mL and a lower limit of 8.023 pg/mL. The biosensor had excellent repeatability, selectivity, and stability. Moreover, it can provide a new method for the sensitive and rapid detection of other protein molecules in clinical practice.

关键词