2025

Copper and nickel doped carbon dots for rapid and sensitive fluorescent turn-off detection of bilirubin.

Scientific reports Lakshmi Devi A, Sreelakshmi M, Suneesh PV, Satheesh Babu TG
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Carbon dots doped with metals and non-metals have gained much popularity due to the enhancement in their optical and electronic properties. In this study, polyethyleneimine-functionalized transition metal (nickel or copper) doped carbon dots (CD, NiCD and CuCD) were synthesized through hydrothermal method. The carbon dots exhibited a blue fluorescence at 470 nm when excited at 350 nm. The as-synthesized carbon dots were utilised for the fluorimetric detection of bilirubin in the range 0.5 µM - 280 µM, with CuCD exhibiting the highest sensitivity of 155.38 a.u/log µM in the concentration range 0.5 to 10 µM and 84.01 a.u/ log µM in the concentration range 10 to 280 µM. CuCD also exhibited the lowest limit of detection of 0.0907 µM and the lowest limit of quantification of 0.3023 µM. All the carbon dots showed negligible interference in the presence of biomolecules and metal ions present in human serum implying the remarkable selectivity of the method to bilirubin detection. Further, the carbon dots were successfully tested for their real-time application in human serum using bilirubin-spiked serum samples.

关键词