组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
We present a lateral flow aptasensor for the visual detection of alpha-fetoprotein (AFP) in human serum. Leveraging the precise molecular recognition capabilities of aptamers and the distinct optical features of gold nanoparticles, a model system utilizing AFP as the target analyte, along with a pair of aptamer probes, is implemented to establish proof-of-concept on standard lateral flow test strips. It is the first report of an antibody-free lateral flow assay using aptamers as recognition probes for the detection of AFP. The analysis circumvents the numerous incubation and washing steps that are typically involved in most current aptamer-based protein assays. Qualitative analysis involves observing color changes in the test area, while quantitative data are obtained by measuring the optical response in the test zone using a portable strip reader. The biosensor exhibits a linear detection range for AFP concentrations between 10 and 100 ng/mL, with a minimum detection limit of 10 ng/mL. Additionally, it has been successfully applied to detect AFP in human serum samples. The use of aptamer-functionalized gold nanoparticle probes in a lateral flow assay offers great promise for point-of-care applications and fast, on-site detection.