组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
We demonstrate a flexible electrochemical biosensor for non-enzymatic glucose detection under different bending conditions. The novel flexible glucose sensor consists of a Cu nanoparticle (NP)/laser-induced graphene fiber (LIGF)/porous laser-induced graphene (LIG) network structure on a polyimide film. The bare LIGF/LIG electrode fabricated using an 8.9 W laser power shows a measured sheet resistance and thickness of 6.8 Ω/□ and ~420 μm, respectively. In addition, a conventional Cu NP electroplating method is used to fabricate a Cu/LIGF/LIG electrode-based glucose sensor that shows excellent glucose detection characteristics, including a sensitivity of 1438.8 µA/mM∙cm2, a limit of detection (LOD) of 124 nM, and a broad linear range at an applied potential of +600 mV. Significantly, the Cu/LIGF/LIG electrode-based glucose sensor exhibits a relatively high sensitivity, low LOD, good linear detection range, and long-term stability at bending angles of 0°, 45°, 90°, 135°, and 180°.