2025

Electrochemical Detection of Prostate Cancer-Associated miRNA-141 Using a Low-Cost Disposable Biosensor.

Biosensors Hunt A, Slaughter G
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Prostate cancer is the second leading cause of cancer-related deaths among men in the United States. The early detection of aggressive forms is critical. Current diagnostic methods, including PSA testing and biopsies, are invasive and often yield false results. MicroRNA-141 (miRNA-141) has emerged as a promising non-invasive biomarker due to its elevated levels in the urine of patients with metastatic prostate cancer. Here, a low-cost, paper-based electrochemical biosensor for the sensitive detection of miRNA-141 in synthetic urine is reported. The device employs inkjet-printed gold electrodes on photopaper, functionalized with thiolated single-stranded DNA-141 capture probes for specific target recognition. The biosensor achieves a sensitivity of 78.66 fM µA-1 cm-2 and a linear detection range of 1 fM to 100 nM, encompassing clinically relevant concentrations of miRNA-141 found in patients with metastatic prostate cancer. A low limit of detection of 2.15 fM, strong selectivity against non-target sequences, and a rapid response time of 15 min further highlight the diagnostic potential of the device. This platform represents a significant advancement in the development of point-of-care diagnostic tools for prostate cancer and is readily adaptable for detecting other disease-specific miRNAs through simple probe modification. As such, it holds broad promise for accessible, early-stage cancer detection and longitudinal disease monitoring in diverse clinical settings.

关键词