A Proof-of-Concept Study on Bioelectric-Based Biosensing for Prostate-Specific Antigen Detection in Serum Samples.

Biosensors Giannakos G, Marka S, Georgoulia K, Kintzios S, Moschopoulou G
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Prostate cancer is among the most prevalent malignancies in men worldwide, underscoring the need for early and accurate diagnostic tools. This study presents a proof-of-concept and pilot clinical validation of a novel bioelectric impedance-based biosensor for the detection of prostate-specific antigen (PSA) in human serum. The system integrates Molecular Identification through Membrane Engineering (MIME) with the xCELLigence real-time cell analysis platform, employing Vero cells electroinserted with anti-PSA antibodies. Optimization experiments identified 15,000 cells/well as the optimal configuration for impedance response. The biosensor exhibited specific, concentration-dependent changes in impedance upon exposure to PSA standard solutions and demonstrated significant differentiation between PSA-positive and PSA-negative human serum samples relative to the clinical threshold of 4 ng/mL. The biosensor offered rapid results within one minute, unlike standard immunoradiometric assay (IRMA), while showing strong diagnostic agreement. The system's specificity, sensitivity, and reproducibility support its potential for integration into point-of-care screening workflows. This bioelectric assay represents one of the fastest PSA detection approaches reported to date and offers a promising solution for reducing overdiagnosis while improving clinical decision-making and patient outcomes.