2025

A Highly Sensitive Silicon Nanowire Array Field Effect Transistor Biosensor for Detecting HBV-DNA and AFP.

Sensors (Basel, Switzerland) Sun P, Liu M, Zhang Y, Liu C, Yang X
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Chronic hepatitis B poses a serious threat to human health and life, and early diagnosis is essential to improving patient cure rates. Hepatitis B virus (HBV) and Alpha-fetoprotein (AFP) are two key biomarkers for diagnosing chronic hepatitis B. In this study, we propose a silicon nanowire array field effect transistor (SiNW-array FET) biosensor that enables highly sensitive, real-time, and low-cost joint detection of both HBV and AFP. The SiNW-array FET is fabricated using traditional micro-nano fabrication techniques such as self-limiting oxidation and anisotropic etching, and its morphology and electrical properties were tested. The results show that the diameters of the fabricated silicon nanowires (SiNWs) are uniform and the SiNW-array FET exhibits a strong output signal and high signal-to-noise ratio. Through specific chemical modification on the surface of SiNWs, the SiNW-array FET is highly sensitive and specific to HBV-DNA fragments and AFP, with ultralow detection limits of 0.1 fM (HBV-DNA) and 0.1 fg/mL (AFP). The detection curve of the SiNW-array FET exhibits good linearity within the HBV-DNA concentration range of 0.1 fM to 100 pM and AFP concentration range of 0.1 fg/mL to 1000 pg/mL. More importantly, the device could also detect HBV-DNA successfully in serum samples, laying a solid foundation for the highly sensitive clinical detection of chronic hepatitis B.