2025

High-performance hexagonal cavity-based plasmonic MIM waveguide biosensor for sensitive and label-free detection of carcinoembryonic antigen (CEA).

Scientific reports Bahador H, Izadiyar H, Abdoli Z, Nilghaz A
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

In this research, a plasmonic metal-insulator-metal (MIM) biosensor is proposed and numerically analyzed for label-free and highly sensitive refractive index detection. The design incorporates a hybrid configuration consisting of a hexagonal resonant cavity coupled with a narrow rectangular slit along its lower boundary. This particular arrangement, which has not been previously reported, enables the excitation of two distinct resonance modes that contribute to improved spectral resolution and sensing precision relative to conventional MIM structures. Finite-difference time-domain (FDTD) simulations were employed to investigate the optical response and optimize the geometrical parameters of the device. The optimized sensor demonstrates strong electromagnetic field confinement, achieving a sensitivity of 770.27 nm/RIU and a figure of merit (FoM) of 160.47, both of which surpass values reported in related studies. Owing to its compact geometry and compatibility with nanoimprint lithography, the proposed design can be fabricated with high practicality. These findings suggest that the device offers an efficient and scalable platform for real-time, label-free detection of carcinoembryonic antigen (CEA) and other biomarkers relevant to early-stage disease diagnostics.