组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
Early diagnosis of Alzheimer's disease (AD) remains challenging due to the extremely low concentration of relevant biomarkers and the limited sensitivity of conventional detection techniques. In this study, we present a carbon nanotube field-effect transistor (CNT-FET) immunosensor for label-free detection of phosphorylated tau at threonine 217 (p-tau217). The device employs a Y2O3/HfO2 dielectric layer and gold nanoparticles (AuNPs) to improve biofunctionalization, with anti-p-tau217 antibodies immobilized on the CNT channels. In phosphate-buffered saline (PBS), the sensor exhibited a linear response over a concentration range of 3 fM to 30 pM (R2 = 0.973) and achieved a limit of detection (LOD) of 1.66 fM. The device demonstrated high selectivity, with a normalized signal response (NSR) for p-tau217 that was 5-6 times higher than for human serum albumin (HSA) and p-tau231, even at 1000-fold higher concentrations of these interferents. The sensor exhibited reproducibility with a relative standard deviation (RSD) of 4.8% (n = 9) and storage stability with only a 10% decrease in signal after 7 days at 4 °C. Mechanistic analysis indicated that the net positive charge and structural flexibility of the p-tau217 peptide led to a reduction in drain current upon binding, consistent with electrostatic gating effects in p-type CNT-FETs. Current limitations include the absence of standardized p-tau217 reference materials. Future work will focus on validation with clinical samples. This CNT-FET platform enables rapid, minimally invasive detection of p-tau217 and holds strong potential for integration into clinical workflows to facilitate early AD diagnosis.