2025

A Wearable Infrared Sensor for Detecting Non-ST Segment Elevation Acute Coronary Syndromes.

medRxiv : the preprint server for health sciences Sengupta PP, Jamthikar AD, Yanamala N, Maganti K, Titus J, Bhavnani SP, Daniels L, Peacock WF, Sengupta S, iSENSE-ACS multicenter study investigators
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Non-ST-segment elevation acute coronary syndrome (NSTE-ACS) is conventionally diagnosed using electrocardiography and serial blood biomarker measurements. We investigated a non-invasive, bloodless, and electrode-free diagnostic strategy using a wrist-worn infrared spectrophotometric biosensor (Infrasensor). In a prospective, multicenter study of 595 patients with suspected NSTE-ACS enrolled across 13 sites in two countries, participants were stratified into five analytical cohorts. With 200 multi-ethnic controls and a leave-one-cohort-out external validation, a machine learning model detected high-grade coronary obstruction with an area under the receiver operating characteristic curve (AUC) of 0.87 (95% CI: 0.84-0.90), 90% specificity, and 84% positive predictive value-surpassing standard risk scores. A secondary model predicted freedom from NSTE-ACS and adverse outcomes over 30 days with an AUC of 0.89 (95% CI: 0.87-0.92), 99% sensitivity, and 96% negative predictive value. These findings demonstrate the potential of the Infrasensor as a rapid, scalable point-of-care tool for early risk stratification in NSTE-ACS.

关键词